All the equipment needed to build up your own Calibration Laboratory is available as off-the-shelf products from Norsonic.
Acoustical and electrical calibration of sound measuring instruments, measuring microphones and sound calibrators in accordance with the applicable traceable standards.
Periodic Verification of Sound Calibrators
- Sound pressure level corrected for pressure, temperature and humidity
- Short term stability expressed in dB (option)
- Output frequency in Hz
- Total distortion in % (option)
Periodic Verification of Sound Level Meters
The Periodic Verification of Sound Level Meters can be performed in accordance with several national and international standards. This includes measurements of:
- Accuracy of input range selector
- Level linearity
- Spectral weighting networks A, B, C, Z, Lin and Flat
- F, S, and I time constants. Pulse measurement capability
- RMS detector capability
- Overload detector
- Inherent noise using microphone equivalent or acousti-cally insulated test vessel
- Integrating averaging functions Leq and SEL
- Noise dose
- 1/1 and 1/3 octave band filters (option)
Features
- Frequency response calibration of microphones using the electrostatic actuator (with swept-sine) or acoustical method
- Sensitivity calibration of microphones using the insert voltage method ensures a high degree of accuracy
- Fast and accurate calibration of sound calibrators, pistonphones and associated barometers in accordance with IEC 60942 (2003 and 2017)
- Sound level meter calibration in accordance with IEC 61672-3 (2006 and 2013), IEC 60651:1979, IEC 60804:2000, DIN 45657:2014, ANSI S1.4 (2014 and older), ANSI S1.43-1997 and BS 7580:1997
- Fast and accurate calibration of sound measuring instruments using three different interface modes; manual; semi-automated and fully automated
- Test of fractional octave filters in accordance with IEC 61260-3:2016
- 1260-3:2016. • Full test report generated
- Built-in self test features (requires voltmeter)
- Easy to set up, includes all accessories needed
Downloads:
Product data Norsonic Nor1525 calibration lab equipment